3-42
(5) Pachymetry map
Thinnest:
The corneal thickness [μm] at the thinnest section
and the position (X/Y coordinates) from the corneal
vertex [mm]
(6) Instantaneous Power [Posterior] map
Steepest@6:
The minimum value [mm] of Instantaneous Power
[Posterior] within the range of Φ6 mm and the
position [mm] (X/Y coordinates) from the corneal
vertex
i) IOL Screening
When using results obtained from this instrument to
select intraocular lenses, also carefully consider other
examination methods.
When a corneal shape map is used for diagnosis, a
thorough review is necessary, such as performing
measurements multiple times and performing other
tests. Note that analysis errors may occur due to light
blockage by eyelashes and eyelids, misalignment of
the center due to unstable sight, conditions of corneal
disorder and/or related secretions, etc.
This application is for an examination before cataract
surgery. Function for 25 cones only.
Data related to entry of IOL formulae and selection of
spherical/aspherical/ToricIOL is displayed to allow you to
select the appropriate IOL.