26. SPECIFICATIONS
Measuring range
*9
(Using the following reflective prism/reflective sheet target during
normal atmospheric conditions
*1
/
*2
is good atmospheric conditions)
360° Prism ATP1/ATP1S:
*2
1.3 to 1,000 m (3,280ft)
*3
Prism-5: 1.3 to 500m (1,640ft)
*7
Prism-2 X 1: 1.3 to 5,000m (16,400ft)
*7
(to 6,000m) (19,680ft)
*7
Prism-2 X 3: to 8,000 m (26,240ft)
*7
(10,000m) (32,800ft)
*7
Reflective sheet RS90N-K: 1.3 to 500m (1,640ft)
*4
Reflective sheet RS50N-K: 1.3 to 300m (980ft)
*4
Reflective sheet RS10N-K: 1.3 to 100m (320ft)
*4
Reflectorless (White) : 0.3 to 800m (2,620ft)
*5
(to 1,000m) (3,280ft)
*6
Prism (tracking)
*3
:1.3 to 1,000m (3,280ft)
Refrective sheet target (tracking)
*4
:1.3 to 350m (1,140ft)
Reflectorless (White) (tracking)
*5
:0.3 to 300m (980 ft)
Minimum display
Fine measurement:
0.0001m (0.001ft/ 1/16 inch)/0.001 m (0.005ft/ 1/8 inch)
Rapid
measurement:
0.001 m (0.005ft/ 1/8 inch)
Tracking measurement: 0.01 m (0.1ft/ 1/2 inch)
Maximum slope distance display (Except for tracking)
12,000.000m (39,370ft) (using prism or reflective sheet target)
1,200.000m (3,930ft) (Reflectorless)
Distance unit m/ft/inch (selectable)
Accuracy (D: measurement distance; Unit: mm) (Under normal atmospheric conditions
*1
)
*7, *9
(Using prism or using 360° Prism ATP1/ATP1S
*3
)
Fine measurement: ±(1.5 + 2 ppm X D) mm
Rapid measurement: ±(5 + 2 ppm X D) mm
(Using reflective sheet target)
*4
Fine measurement: ±(2 + 2 ppm X D) mm
Rapid measurement: ±(5 + 2 ppm X D) mm
(Reflectorless (White))
*5
Fine measurement:
±(2 + 2ppm X D) mm (0.3 to 200m)
*8
±(5 + 10ppm X D) mm (over 200 to 350m)
±(10 + 10ppm X D) mm (over 350 to 10,00m)
Rapid measurement:
±(6 + 2ppm X D) mm (0.3 to 200m)
±(8 + 10ppm X D) mm (over 200 to 350m)
±(15 + 10ppm X D) mm (over 350 to 10,00m)
(ISO 17123-4 : 2001)
Measurement mode Fine measurement (single/repeat/average)/Rapid measurement
(single/repeat)/Tracking (selectable)
Measuring time (fastest time under good atmospheric conditions
*2
, no compensation,
EDM ALC at appropriate setting, slope distance)
Fine measurement: less than 1.5 sec + every 0.9 sec or less
Rapid measurement: less than 1.3 sec + every 0.6 sec or less
Tracking measurement: less than 1.3 sec + every 0.4 sec or less

Table of Contents