3 Length/Strength Calibration
The length/strength module has one (1) calibration type available:
1. LS Cotton
This can be selected to Check Calibration. The L/S Cotton
calibration type allows the operator to:
Setup Tolerances
Setup References
View Calibration Results
3.1 L/S Cotton Calibration
1. Highlight Length+Strength in the Module window.
2. Highlight LS Cotton in the Calibration Type window.
3.1.1 Setup L/S Calibration Tolerances
Range tolerances ensure that the slope is not distorted by data that are
at opposite tolerance limits. For example, if the measured value for one
of the calibration cottons is at the upper limit of the tolerance, the range
value ensures that the corresponding value for the other calibration
cotton cannot be at the lower limit of the tolerance. A typical range
tolerance is 1.5 times the corresponding tolerance.