USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc. Chapter 5
(3/08) Calibration - Page 5-16
4 Micronaire Calibration
The calibration procedure for micronaire involves the regulation of
airflow through the micronaire chamber when two calibration cottons
(representing low and high micronaire values) are tested.
Figure 5-7 Calibration Main Menu
If references need to be edited or updated, do the following:
1. Highlight Micronaire in the Module window, and then highlight
Mic Cotton in the Calibration Type window.
2. Click Setup References button.
3. Enter a valid password and click OK.
4. Follow instructions below.