EasyManua.ls Logo

USTER HVI 1000 - Check Calibration

USTER HVI 1000
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc. Chapter 5
(3/08) Calibration - Page 5-10
3.1.3 Check Calibration
Click the Check Calibration button to launch the CALIBRATION MENU.
Click Start to perform a calibration check, or to begin calibration.
Figure 5-5 L/S Check Calibration Menu
If the calibration check passes, an L/S Calibration is not required. To
perform a re-calibration anyway, click Force a Re-Calc.
If the calibration check fails, you must then perform L/S calibration.
When calibration procedures are complete results are saved
automatically as calibration passes. You can then select to view or print
a Report. Click Finish to return to the C
ALIBRATION MAIN MENU.

Table of Contents

Related product manuals