Uster Technologies, Inc. Chapter 5
(3/08) Calibration - Page 5-19
Figure 5-10: Micronaire Chamber and Door Sensor
To rerun the calibration again, release the door sensor and press it
again.
Once calibration is complete, the software will indicate whether the
calibration is within tolerance by reporting Pass/Fail. If the calibration
constants are out of tolerance, the calibration procedure will have to be
performed again. If this condition persists, the micronaire chamber may
need to be resized by an Uster service technician.
You can now click Report to view or print the MIC Calibration Report.
Click Finish to return to the C
ALIBRATION MAIN MENU.