15.11.2.1 No flow
Probable causes Action
1
No fresh water supply.
(low discharge pressure < 2 bar
during flushing).
(< 0.2 MPa during flushing).
Check if the 3-phase power supply is switched on.
Check the fuses of the 3-phase power supply. the
water supply to the pump.
2
No sample water supply.
(low discharge pressure < 2 bar
during sampling).
(< 0.2 MPa during sampling).
Check if the cargo/stripping pump is ON.
Check if the discharge line is filled with water.
Check if the manual valve of the sample probe is open.
Check if the pneumatic valve in the sample line is open.
Check if the manual valve in the sample inlet of the skid
assembly is open.
3
Blockage in outlet line.
(high discharge pressure > 10
bar).
(>1 MPa).
Check if the manual valve in the sample outlet of the
skid assembly is open.
Check if the manual valve on the discharge to the slop
tank is open.
4
Flowrate through the monitor too
low
Check the flowrate through the monitor.
Read the skid flow from the display during FORWARD
FLUSH or SMAPLE mode and ensure the flow is
between
450 - 550 l/h.
The flowrate can be adjusted by means of the needle
valve on the outlet of measurement cell.
5
Malfunction of the flow sensor.
Check the electrical connections on the flow sensor.
Check the connections between the flow sensor and
the EPU.
Check the function of the flow sensor.
NOTE:
During commissioning the flowrate through the monitor is set to between 450 and 550
l/hour. The “no-flow” alarm is activated at flows less than 240 litres per hour.
NOTE:
During back flush, the “no flow” alarm is inhibited. The flow alarm comes into operation
at the start of the forward flush cycle.