57
7.2.1.5.3 Calibration flush mode (menu 1.5.3.0)
After the <FLUSH> key in the setup mode (menu 1.5) is pressed, the system prepares to go into
sampling mode by initiating a calibration flush sequence.
The calibration flush sequence is used to clean the system prior to sampling and to check the
condition of the detector cell and the system. This sequence consists of approximately 2 minute
backward flush, 12 seconds window wash, 40 seconds zero check and a total of 4 minute forward
flush and calibration (total of approximately 7 minutes).
After completion of the calibration flush sequence and if no failures are detected the system will
automatically go into idle mode.
Figure 47
During the flush sequence no keys are available on the display.
When the sequence has finished, the keys will reappear and the system will be in idle mode (menu
1.5.3)
Figure 48
CONTROL Sample point: 1 22-11-11
Calibration flush Oil type: 5 10:38
Concentration A 0 PPM discharge
Flow A 0 m3/h dirty
Skid Flow 0 l/h com: C
Speed A 10.0 kts pos: C
Discharge rate 0.0 l/nm segregated
Total oil 0.0 l com: C
Latitude 5321.756N pos: C
Longitude 00711.854E
Backflush
EPU: OK
free: 1927296 kB
CONTROL Sample point: 1 22-11-11
idle Oil type: 5 10:38
Concentration A 0 PPM discharge
Flow A 0 m3/h dirty
Skid Flow 0 l/h com: C
Speed A 10.0 kts pos: C
Discharge rate 0.0 l/nm segregated
Total oil 0.0 l com: C
Latitude 5321.756N pos: C
Longitude 00711.854E
EPU: OK
free: 1927296 kB
SAMPLE? CHANGE SELFTESTSTOP