15.4.9 Discharge rate failure
Probable causes Action
1
Discharge rate to high.
Check the overboard discharge rate and reduce it to
clear the alarm.
Also consult section 7.5
15.4.10 Total discharge oil limit alarm
Probable causes Action
1
The maximum oil limit is
reached
STOP the discharge operation immediately.
Also consult section 7.5
NOTE:
The maximum amount of oil for this voyage is discharged. the total oil may only be
reset at the start of a new ballast voyage.
15.5 GPS PROBLEMS
15.5.1 No GPS signal visible
Probable causes Action
1
Wrong GPS settings
Check that GPS setting are:
GPS: NMEA 0183
Baud rate: 4800
Data bits: 8
Parity: none
Stop bits: 1
2
Internal GPS failure
Disconnect the GPS signal cable from socket X14.
Connect terminals 1 - 3 from socket X14.
Connect terminals 2 - 4 from socket X14.
A GPS NMEA string will be simulated.
If no GPS coordinates are showing there is an internal
failure.