EasyManua.ls Logo

VAF instruments OILCON MARK 6M - 3.4 SKID; 3.5 SAMPLE PUMP

VAF instruments OILCON MARK 6M
175 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
3.4 SKID
Sample flowrate between 450 and 550 l/h
Sample inlet pressure 3 bar (nom.), 6 bar (max.) (0.3 MPa nom., 0.6 MPa max.)
Fresh water supply 3 bar (nom.), 6 bar (max.) (0.3 MPa nom., 0.6 MPa max.)
average consumption 0,13 l/min
max. consumption 8 l/min
Water temperature range 10 °C – 65°C, in accordance with IMO Resolution MEPC 108 (49)
Ambient temperature -20 °C – +55 ºC
Humidity range 0 – 95 % RH
Protection class IP 65
Mounting wall mounting, see drawing 0806-1288 for dimensions
Connections
Fresh water and sample water 15 mm tube coupling
Air to valves 6 mm tube coupling
Air to pneumatic pump 8 mm tube coupling
Cable to EPU M20 x 1,5 (cable diameter 10 – 14 mm)
Wire size connectors Min-max 0,2 – 2,5 mm
2
3.5 SAMPLE PUMP
Electrical supply 3-phase 380–420 VAC, 50Hz, 1.1 kW, 2.5 A
or 3-phase 440–480 VAC, 60 Hz, 1.3 kW, 2.5 A
Power consumption 1300 W
Isolation class F, IEC 34-1
Protection class IP 55
Mounting wall mounting, see drawing 0806-1076 for dimensions
Connections
Sample water 15 mm tube coupling
Cable M20 x 1,5 (cable diameter 10 – 14 mm)
Wire size connectors Min-max 0,2 – 2,5 mm
2

Table of Contents