OPERATING MANUAL_______________________________________________________________
36_____________________________________________________________________M210492EN-A
With some commands (e.g. calibration) the figures are changed using
the arrow keys. When an arrow key is pressed continuously for a
while, the numbers start changing at an increasing rate.
Auto-calibration
NOTE
Normally auto-calibration parameters do not have to be changed.
Select More in the main menu and then again More in the second
More menu. Select Dry cal and then Settings.
The interval parameter defines the frequency of auto-calibration
cycles. When the transmitter is turned on, the first auto-calibration
takes place after an hour unless the frequency has been set to less
than an hour. After the first auto-calibration, the set frequency is
activated and the auto-calibration takes place e.g. every six hours.
If the setting is correct, press ENT.
If the setting needs to be changed, press CL and change the setting
with arrow keys; acknowledge the setting with ENT. If the setting
is changed, it becomes valid only after the next auto-calibration
has been completed. If you wish to activate it immediately reset
the transmitter or turn it off.
This parameter defines the maximum allowed change of the
dewpoint value during the pre-defined dTdp time. If the change in
dewpoint value exceeds the limit, the calibration is not started. The
auto-calibration is done only after process is stabilized. The
transmitter retries untill the calibration is succesfully complited.
If the setting is correct, press ENT.
If the setting needs to be changed, press CL and change the value
with arrow keys; acknowledge the setting with ENT.