CHAPTER 5________________________________________________________________ COMMANDS
VAISALA _________________________________________________________________________ 37
This parameter defines the measuring time of the change in
dewpoint prior to the auto-calibration (see the parameter Max
dTdp).
If the setting is correct, press ENT.
If the setting needs to be changed, press CL and change the value
with arrow keys. Acknowledge the setting with ENT.
This parameter defines the maximum correction (%RH) the
transmitter does during each auto-calibration cycle. If the
correction exeeds the limit, the calibration is ignored. If the setting
is correct, press ENT.
If the setting needs to be changed, press CL and change the value
with arrow keys; acknowledge the setting with ENT.
This parameter defines the time the output values prior to the auto-
calibration are frozen after the calibration. The time is for sensor
temperature stabilization.
If the setting is correct, press ENT.
If the setting needs to be changed, press CL and change the value
with arrow keys; acknowledge the setting with ENT.
NOTE
Auto-calibration takes place only when the security lock jumper is
connected. During auto-calibration, the reading on the display is
frozen.