EasyManua.ls Logo

Vaisala DMP248 - Mounting; overview

Vaisala DMP248
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 2_______________________________________________________ PRODUCT DESCRIPTION
VAISALA __________________________________________________________________________ 5
The function Pws(T) is well known from literature. Thus, it is
possible to determine the offset (RH
0
) if measurements are made at
two or (preferably) more temperatures assuming a constant Pw
during the process.
30°C
20°C
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0 0.005 0.01 0.015 0.02 0.025 0.03 0.035 0.04 0.045
1/Pws(T)
RHout(%)
FIGURE 2-1 Device output during hypothetical offset calibration
cycle.
For example, for a transmitter with a -0.2%RH offset error is made 11
measurements at temperatures 20...30 °C. The frostpoint is assumed to
stay constant at -40 °C and that the 'Gain' is nominal 100%.
As the temperature increases the RH drops. Since the response is
linear, a straight line is fitted through the data points. The line
intercepts the y-axis at RH
0
as shown in FIGURE 2-1. Now the RH
0
is
known and it can be subtracted from the indicated value 0.35% RH to
get the correct value of 0.55% RH.
FIGURE 2-2 shows the DRYCAP
sensor as mounted on a DMP248
probehead. Through a combination of the polymer sensor and Pt 100
sensor, the DRYCAP
will accurately measure the water vapor
pressure used in determining low dewpoints.
During auto-calibration
the Pt 100 element is used to first heat and then measure the
temperature of the sensor while cooling back to ambient temperature.
A complete cycle of auto-calibration takes 60...70 seconds. When the
auto-calibration cycle is active, the transmitter locks the output values
to those measured prior to auto-calibration.

Table of Contents

Other manuals for Vaisala DMP248

Related product manuals