Recommendations
Continuous use over high operating voltage may cause heating. To conserve power and
minimize the warming of the probe, use the lowest operating voltage in the allowed
range.
Using low impedance loads on the signal outputs increase the current consumption by up
to 0.5 mA. High impedance loads are recommended to minimize warming of the probe.
Frequent interrogation of the probe using the RS-485 interface will also increase current
consumption from the average value. More frequent interrogation than once per second is
not recommended.
HMP60 and HMP110 Series User Guide M211060EN-J
28