EasyManuals Logo
Home>Vaisala>Measuring Instruments>HMP110REF

Vaisala HMP110REF User Manual

Vaisala HMP110REF
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
9. Technical data
9.1 HMP60 specifications
Table 17 HMP60 measurement performance
Property Description/Value
Relative humidity
Measurement range 0 … 100 %RH
Typical accuracy:
at 0 … +40 °C (+32 … +140 °F) ±3 %RH (0 … 90 %RH)
±5 %RH (90 … 100 %RH)
at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F
and +104 … +140 °F)
±5 %RH (0 … 90 %RH)
±7 %RH (90 … 100 %RH)
Humidity sensor Vaisala INTERCAPâ
Temperature
Measurement range −40 … +60 °C (−40 … +140 °F)
Accuracy:
at +10 … +30 °C (+50 … +86 °F)
at −40 … +10 and +30 … +60 °C (−40 … +50 and
+86 … +140 °F)
±0.5 °C (±32.9 °F)
±0.6 °C (±33.08 °F)
Dew point temperature
Measurement range −40 … +60 °C (−40 … +140 °F)
Typical accuracy:
at +32 … +40 °C (0 … +140 °F)
when dew point depression
1)
< +15 °C (< +59 °F)
±2 °C (±35.6 °F)
at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F
and +104 … +140 °F) when dew point depression
< +10 °C (< +50 °F)
±3 °C (±37.4 °F)
Analog outputs
Accuracy at +20 °C (+68 °F) ±0.2 % of FS
Temperature dependence ±0.01 % of FS/°C (±0.006 % of FS/°F)
1) Dew point depression = ambient temperature − dew point.
HMP60 and HMP110 Series User Guide M211060EN-J
70

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala HMP110REF and is the answer not in the manual?

Vaisala HMP110REF Specifications

General IconGeneral
BrandVaisala
ModelHMP110REF
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals