User's Guide _______________________________________________________________________
50 ___________________________________________________________________ M211060EN-H
Clear Adjustment of T Measurement
Use the CTCLR command to clear the adjustment of temperature
measurement that has been done using the CT command.
CTCLR<cr>
Example:
ctclr
OK
View User Adjustment Parameters
Use the L command to view the current user adjustment parameters. This
command is useful for checking the currently applied customer
calibration.
L<cr>
The output values are as follows:
- Cp offset and gain: capacitance, calibrated using the CRH command
- T offset and gain: calibrated using the CT command
Example (shows default values, no user calibration done):
l
Cp offset : 0.00000000E+00
Cp gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00