EasyManua.ls Logo

Vaisala HMP110REF

Vaisala HMP110REF
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
48 ___________________________________________________________________ M211060EN-H
Calibration Commands
Calibrate Humidity Measurement
Use the CRH command to perform a one-point or two-point correction to
the capacitance measurement of the probe. This command changes the
offset and/or gain of the humidity measurement, depending on the
calibration and reference:
- one-point calibration with a single < 50 %RH reference will adjust the
offset of the capacitance measurement
- one-point calibration with a single > 50 %RH reference will adjust the
gain of the capacitance measurement
- two-point calibration will adjust both offset and gain. The first point
requires a < 50 %RH humidity reference, the second point must be
> 50 %RH. There must also be at least 30 percentage point difference
between the references.
CRH [reference]<cr>
This command is not available on the HMP110T.
When performing a one-point calibration, you need to place the probe in
the reference humidity and wait for 20 – 40 minutes for the humidity to
stabilize. To apply the adjustment, enter the CRH command with the
reference %RH as a parameter.
Example: one-point calibration (LiCl reference, 11 %RH):
crh 11
OK
Example: one-point calibration with NaCl reference (75 %RH):
crh 75
OK
Giving the command without parameters starts the two-point calibration.
Remember to allow the humidity to stabilize for 20 - 40 minutes after
changing the reference.
Example: two-point calibration with LiCl (11 %RH) and NaCL
(75 %RH) references:
crh
RH : 11.2684 1. ref ? 11
Press any key when ready ...
RH : 75.0612 2. ref ? 75
OK

Table of Contents

Related product manuals