Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 41
View Calibration Information
Use the CDATE command to view the calibration date and CTEXT to
view the calibration info text. Date format for CDATE is
YYYYMMDD.
CDATE<cr>
CTEXT<cr>
Examples:
cdate
Cal. date : 20150109
ctext
Cal. info : VAISALA/HEL
Enter Calibration Information
To enter the calibration date, use the CDATE command (cdate
[YYYYMMDD]
). To enter a text string with information about the
calibration, use the CTEXT command (ctext [string]).
CDATE [yyyymmdd] <cr>
CTEXT [text string] <cr>
Example:
cdate 20151125
Cal. date : 20151125
ctext Calibrated in Room 1
Cal. info : Calibrated in Room 1
View Order Code
Use the CODE command to view the order code that has been stored in
the probe. This command is useful if you need to order a new probe with
the same options.
CODE<cr>
Example:
code
Order code : A12A0A2B0