Transmitter error code in MI70 Corresponding error
8
hex
Capacitance reference error. [47]
10
hex
Ambient temperature out of range. [48]
20
hex
Firmware checksum mismatch. [49]
40
hex
Device settings corrupted. [50]
80
hex
Additional configuration settings corrupted. [51]
100
hex
Sensor coecients corrupted. [52]
200
hex
Main configuration settings corrupted. [53]
800
hex
Supply voltage out of range. [55]
2000
hex
Non-volatile memory read/write failure. [57]
4000
hex
Calibration certificate checksum mismatch. [58]
8.5 Unknown serial settings
You can check the current serial settings of your probe by connecting to the serial line, or with
Vaisala Insight software.
More information
Vaisala Insight software (page 31)
Accessing serial line command interface from analog or Modbus mode (page 36)
Chapter 8 – Troubleshooting
69