Transmitter error code in MI70 Corresponding error
8
hex
Capacitance reference error. [47]
10
hex
Ambient temperature out of range. [48]
20
hex
Firmware checksum mismatch. [49]
40
hex
Device settings corrupted. [50]
80
hex
Additional configuration settings corrupted. [51]
100
hex
Sensor coecients corrupted. [52]
200
hex
Main configuration settings corrupted. [53]
800
hex
Supply voltage out of range. [55]
2000
hex
Non-volatile memory read/write failure. [57]
4000
hex
Calibration certificate checksum mismatch. [58]
8.5 Unknown serial settings
You can check the current serial settings of your probe by connecting to the serial line, or with
Vaisala Insight software.
More information
‣
Vaisala Insight software (page 31)
‣
Accessing serial line command interface from analog or Modbus mode (page 36)
Chapter 8 – Troubleshooting
69