Example (probe in STOP mode, no address needed):
send
T= 22.7 'C RH= 20.0 %RH Td= -1.5 'C
Example (probe in POLL mode, with address 10):
send 10
T= 22.8 'C RH= 20.1 %RH Td= -1.3 'C
6.8 Configuring serial line operation
6.8.1 Set serial line settings
Use the SERI command to show or set the serial line settings. The new settings will be taken
into use when the probe is reset or powered up.
SERI [b p d s]
Syntax
Description
b
baud rate (300, 600, 1200, 2400, 4800, 9600, 19200, 38400,
57600)
p
parity (n = none, e = even, o = odd)
d
data bits (7 or 8)
s
stop bits (1 or 2)
Example (shows default settings):
seri
Baud P D S : 19200 N 8 1
6.8.2 Set serial interface mode
Use the SMODE command to set the operation mode of the serial interface. The new mode is
applied when probe is reset.
SMODE [xxx]<cr>
Syntax
Description
xxx
Operation mode of the serial interface.
HMP60 and HMP110 Series User Guide M211060EN-J
42