EasyManua.ls Logo

Vaisala HMP110REF

Vaisala HMP110REF
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
30 ___________________________________________________________________ M211060EN-H
Power Supply Requirements
The operating voltage for the HMP60 and HMP110 series probes must be
in the following range:
Table 7 Operating Voltage Ranges
HMP60 / HMP63 / HMP110 analog /
HMP113 / HMP110T
HMP110REF / HMP110 digital
5 ... 28 VDC (V
out
0 ...1 / 0 ... 2.5 V)
5 ... 28 VDC
8 ... 28 VDC (V
out
0 ... 5 / 1 ... 5 V)
Current consumption is 1 mA on average, which makes the probes well
suited for running on battery power. The maximum peak consumption is
5 mA.
Recommendations
- Continuous use over high operating voltage may cause heating.
To conserve power and minimize the warming of the probe, use the
lowest operating voltage in the allowed range.
- Using low impedance loads on the signal outputs increase the current
consuption by up to 0.5 mA. High impedance loads are recommended
to minimize warming of the probe.
- Frequent interrogation of the probe using the RS-485 interface will
also increase current consumption from the average value. More
frequent interrogation than once per second is not recommended.

Table of Contents

Related product manuals