Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 55
Example (on analog output, extend the maximum RH reading to 120%
and scale the RH output on channel 2 from (0…1 V) 0...100% to (0…1
V) 0…120%):
rhlimit 120
Max. RH % : 120.0 %RH
asel ?
Ch1 T lo : -20.00 'C ?
Ch1 T hi : 80.00 'C ?
Ch2 RH lo : 0.00 %RH ?
Ch2 RH hi : 100.00 %RH ? 120
Display Command List
Use the HELP command to display a list of the currently available
commands.
HELP<cr>
Display the Currently Active Errors
Use the ERRS command to display the currently active error codes. For
troubleshooting active errors, see Table 14 on page 67.
ERRS<cr>
Example (no active errors):
errs
0000h
No errors
Connect to the Probe in POLL Mode
Use the OPEN command to connect to a probe that is in POLL mode.
OPEN [aa]<cr>
where
aa
=
address (0 ... 255)
Example:
open 1
HMP110 1 line opened for operator commands