5. Configure the analog output channel settings:
a. In Output o/on, set the output channel o to edit settings.
b. In Parameter, select the probe and the measurement parameter that controls the
output channel. The unit of the parameter is set automatically.
c. In Scale low end and Scale high end, define the lower and upper ends of the
measurement scale for the chosen output parameter.
d. Set the output channel on.
6. Repeat step 4 … step 5 for all the analog output channels you want to use.
To see the changes you made, select . Then use the left and right arrows at the
bottom of the touchscreen to navigate to the Outputs and relays view.
Chapter 6 – Configuring transmitter using touchscreen display