VAT 651 Series - Initial Operation; Setup Procedure; Logic Interface Configuration

VAT 651 Series
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTALLATION Series
651
38/99
Edition 2014-07-10 605451EC
4.5 Initial operation
4.5.1 Setup procedure
To enable the valve for
pressure control
setup
steps 1 to 7
must
be performed
.
In case position control is required only it’s sufficient to perform steps 1 to 4.
Setup steps Description
1
Power up
Turn on external + 24VDC power supply of valve (and external ±15
VDC for sensor power supply if required).
Refer to chapter «Behavior during power up» for details.
2
Interface
configuration
Refer to chapter «Logic Interface configuration» for details.
3
Valve
configuration
Basic configurations of valve must be adapted according to
application needs. Refer to chapter «Valve configuration» for details.
4
Sensor
configuration
Basic configurations of sensor(s) must be adapted according to
application needs. Refer to chapter «Sensor configuration» for
details.
5
ZERO
Compensation of the sensor offset voltage.
Refer to chapter «ZERO» for details.
6
LEARN
Determination of the vacuum system characteristic to accommodate
the PID controller. Refer to chapter «LEARN adaptive» for details.
7
PRESSURE
CONTROL
COFIGURATION
Accommodation of PID controller to the vacuum system
characteristic.
Refer to chapter: «Pressure Control configuration» for details.
Without «LEARN adaptive» or «Pressure Control configuration» the valve is not able to
run pressure control.
For ease setup (in Local mode) of ‘Interface’, ‘Valve’, ‘Sensor’, ‘Senor ZERO’, ‘LEARN’
and ‘PRESSURE CONTROL COFIGURATION’ it is possible to use the CPA 3.0, The
free download is available on the VAT homepage: http://www.vatvalve.com/customer-
service/informations-and-downloads/control-performance-analyzer
4.5.2 Logic Interface configuration
Interface configuration must be adapted according to application needs.
Default configuration:
OPEN input
CLOSE input
OPEN output
CLOSE output
not inverted not inverted open close
Functionality of digital inputs CLOSE VALVE and OPEN VALVE must be selected.
These may be configured as ‘not inverted‘ or ‘inverted‘. Default is ‘not inverted‘.
LEARN range configuration for remote operation must be selected.
This may either be ‘full range’ or pressure limit according of analog SETPOINT input. Default is ‘full
range‘.