92 Running PowerLine in Direct Control Mode
Vial
Indicates the vial you wish to use for the injection.
To select a vial:
1. Move the cursor to the Vial field.
2. Type the vial number from which to make an injection and press Enter.
Valid vial numbers are 1 to 48 or 1 to 96, depending on the capacity of the carousel
in the autoinjector. You can also enter special vial codes (described below) to
perform special functions.
Special Vial Codes
Vial numbers from 100 to 107 in the Vial field instruct the system to perform the special
functions described in Table 5-1.
Table 5-1 Special Vial Codes
Vial Code Function Description
100 Manual Inj Tells the controller that you will make the injection
with a manual injector or a non-PowerLine
autoinjector.
101 Equilibrate Equilibrates the system at initial conditions for the
specified run time.
102 Purge 410 Purges the Waters 410 detector by placing its flow
path in the purge mode and running initial condi-
tions for the specified run time.
103 Purge Inj Purges the PowerLine autoinjector (Waters
717plus) for 4.5 minutes using the autoinjector
purge routine and initial conditions.
104 Prep Runs blank gradients or reequilibrates a column.
105 Auto Base Instructs the 431/432 detector (Det #3) to execute
its auto-baseline routine.
106 Pos Polarity Instructs the 410 or 431/432 detector (Det #3) to
operate with positive polarity.
107 Neg
Polarity
Instructs the 410 or 431/432 detector (Det #3) to
operate with negative polarity.