EasyManua.ls Logo

Waters 996 - Setting the IEEE-488 Address; Locating the IEEE-488 Address Switches

Waters 996
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18 Installation
1
1.3.2 Setting the IEEE-488 Address
To set the IEEE-488 address for the 996 detector:
1. Use a small screwdriver (or similar device) to set the DIP switches on the detector
rear panel (Figure 1-4
) to the IEEE-488 address of the 996 detector. The address
must be a number from 2 to 29 and must be unique within your network.
Refer to the
Millennium
32
System Installation/Configuration Guide,
Section 2.3.1,
Connecting IEEE-488 Devices, for the correct IEEE-488 DIP switch settings.
Figure 1-4 Locating the IEEE-488 Address Switches
2. To instruct the 996 detector to accept the new IEEE-488 address, power off, then
power on the detector (see Section 1.6,
Startup/Shutdown).
IEE 488 ADDRESS
12345
– OPEN –
TP01457
IEEE-488 Cable
Connection
IEEE-488
Address
Switches

Table of Contents

Related product manuals