996 Detector Optics 45
4
Figure 4-1 Optics Assembly Light Path
Table 4-1
describes the optics assembly components in the 996 detector.
Table 4-1 Optics Assembly Components
Component Function
Lamp and lamp
optics
Focuses light from the deuterium source lamp through a
beamsplitter to the flow cell.
Beamsplitter and
reference diode
Reflects part of the light back to a reference diode, which
measures the intensity of the light emitted by the lamp. The
detector uses this measurement to keep the lamp output constant.
Reference
Diode
Beamsplitter
Assembly
50-mm
Aperture
Lamp and
Lamp Optics
Spectrographic
Mirror and Mask
Shutter
Assembly
Flow Cell
Assembly
Grating
Photodiode
Array