32 Diagnostics and Calibration
2
To prepare for calibration:
1. Set the pump to deliver 1 mL/min of degassed methanol for 10 minutes. If methanol
is not miscible with the previous solvent, flush with a miscible sovent before
switching to methanol.
2. If you have been using buffers, flush with HPLC-quality water at 1 mL/min for 10
minutes, then switch to methanol for 10 minutes.
Note:
Ensure that the solvent is miscible with the previous mobile phase.
For information on performing calibration, refer to the “PDA Calibration Window” topic in
the
Millennium
32
Online Help
Find tab.