Using the column compartments 105
4. Remove the leak sensor from its reservoir, grasping it by its serrations,
and pull upward on it.
5. Use a nonabrasive, lint-free wipe to dry the leak sensor prism.
Caution: To avoid damaging the leak sensor, do not grasp it by the
ribbon cable.
Serrations
TP02891
Prism
Lint-free wipe