Required materials
Gloves: clean, powder-free, chemical-resistant
Replacement active preheater assembly
Required tools
1/4-inch open-end wrench
5/16-inch open-end wrench
To remove the active preheater assembly:
1. Stop any running mobile phase before replacing the active preheater
assembly.
2. After allowing sufficient time for the instrument to cool, open the
column heater, column manager, or auxiliary column manager door.
3. Disconnect the inlet tubing:
Column heater – disconnect the tubing from injector port 6 of the
sample manager.
30-cm column heater – disconnect the tubing from injector port 6 of
the sample manager.
Warning: To prevent burn injuries, set the column temperature to Off,
and then allow the column compartment and its components to cool for
60 minutes before touching them. Monitor the column compartment’s
internal temperature to ensure all components are cool.
Caution: To prevent contaminating system components, wear clean,
chemical-resistant, powder-free gloves when replacing the active
preheater assembly.
Caution: To avoid damaging the active preheater assembly,
turn the column, not the fitting, when installing or removing the
column to prevent the active preheater tubing from turning with the
fitting.