EasyManua.ls Logo

Waters ACQUITY UPLC - Page 23

Waters ACQUITY UPLC
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preparing the column compartments 19
To install the MS-bracket:
1. Stop any currently running mobile phase.
2. Power-off the sample manager.
3. Using the 1/4-inch open-end wrench, disconnect the column inlet, active
preheater, or column stabilizer tubing from injector port 6 of the sample
manager.
4. Disconnect the column-outlet PEEK
®
tubing from the detector inlet.
5. Open the column heater door, and remove the column.
Warning: To prevent burn injuries, set the column temperature to Off,
and then allow the column compartment and its components to cool for
60 minutes before touching them. Monitor the column compartment
internal temperature to ensure all components are cool.
Caution: To prevent contaminating system components, wear clean,
chemical-resistant, powder-free gloves when preparing the column
heater for use with a mass spectrometer.
Column, active preheater, or
column-stabilizer-inlet tubing

Other manuals for Waters ACQUITY UPLC

Related product manuals