Preparing the column compartments 47
CM-A inlet
valve, port 3
Column 1 inlet 12.5 205000730,
APH
205000894,
APH
CM-A inlet
valve, port 4
Column 2 inlet 12.5 205000730,
APH
205000894,
APH
Column 1 outlet CM-A outlet
valve, port 3
19 430002683, SST
tubing,
.004-inch ID
(reorder part
number
700005478)
430003151, SST
tubing,
.003-inch ID
(reorder part
number
700008868)
Column 2 outlet CM-A outlet
valve, port 4
19 430002683, SST
tubing,
.004-inch ID
(reorder part
number
700005478)
430003151, SST
tubing,
.003-inch ID
(reorder part
number
700008868)
Sample
manager
injector valve,
port 6
CM-A inlet
valve, center
port
14.5 430002582, SST
tubing,
.004-inch ID
(reorder part
number
700005476)
430003154, SST
tubing,
.003-inch ID
(reorder part
number
700008869)
Detector inlet CM-A outlet
valve, center
port
14.5 430002681, SST
tubing,
.004-inch ID
(reorder part
number
700005477)
430003158, SST
tubing,
.003-inch ID
(reorder part
number
700008870)
a. Be sure to connect the 430002727 waste tubing to port W (at the 3 o’clock position) of the inlet
valve. Do not connect to the center port.
Two-column configuration plumbing connection details: (Continued)
From To
Length
(inches)
Part number
H-Class I-Class