132 EN MEDUMAT Standard
2
WM 68071 05/2021
6 Operator menu
6.3.8 Presets patient
6-9 Submenu Presets patient
In the Presets patient submenu, you can determine which presets
are assigned to the ventilation parameters of the different patient
groups:
Risk of injury from different alarm presets in the same or
similar devices!
Different alarm presets in the same or similar devices in different
application areas can confuse the user and result in injury to the
patient.
Select the same alarm presets in the same or similar devices.
Risk of injury from deactivated alarms in the additional
functions CPR and RSI!
Through deactivation in the operator menu, no alarms for flow
and CO
2
monitoring are output in the additional functions CPR
and RSI. This can injure the patient.
Notify the user if the alarms in the additional functions CPR
and RSI have been deactivated.
Observe the alarm limits during ventilation in the additional
functions CPR and RSI.

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals