WM 68071 05/2021
MEDUMAT Standard
2
EN 131
6 Operator menu
Suppress maintenance reminder
Activated
Deactivated
When maintenance becomes due, you can
suppress a maintenance reminder within a
maintenance cycle one time for a maximum of
180 days from the date maintenance is due.
The device then behaves as if the maintenance
were not due in terms of the reminders. It is
not possible to reverse suppression of the
maintenance reminder within a maintenance
cycle. After the 180 days have elapsed,
the maintenance reminder becomes active
again until the next service is performed and
cannot be deactivated again.
Even when the maintenance reminder is
suppressed, maintenance is still required
(see "11.2 Intervals", page 223).
Allow filter counter reset
yes
no
You can set whether the user can reset the
hygiene filter at any time in the function
check here.
Filter dust load
100 %
150 %
200 %
Here you can set the load caused by
environmental factors (e.g., dust) for the
hygiene filter.
With an average load (100 %) the hygiene filter
is able to function for approx. 24 ventilation
hours in Air Mix mode or for 6 months.
CO
2
unit
(only with Capnography option)
Vol%
kPa
mmHg
Here you can select which unit of measurement
the CO
2
measurements should be displayed in.
Supply gas
93 % O
2
100 % O
2
Here you can set the type of supply gas.
Trigger setting
3 levels
Units
Here you can set the inspiration and expiration
trigger:
Simple three-level setting
Normal multi-level setting (in units)
Language
The languages
available in the device
are displayed here.
Here you can set the language of the display
texts.
Depending on the status of the device software,
additional languages may be available.
The device shows the languages in your
respective language.
Parameter Possible values Description

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals