WM 68071 05/2021
MEDUMAT Standard
2
EN 135
6 Operator menu
CPR Manual/IPPV
View (only with
activated
Capnography option)
Pressure gauge
Pressure/CO
2
curve
Pressure/etCO
2
trend
Here you can choose between a
pressure gauge view and a curve
view for the CPR mode.
Airway pressure
☑
☐
Here you can determine whether or
not an alarm should be emitted
when airway pressure increases.
CPR Manual
Ventilation pause 2 s - 6 s
Here you can set the time interval for
ventilation between the chest
compressions.
CPR IPPV
MVe /
(only with Flow
measurement + ASB
option)
☑
☐
Here you can determine whether an
alarm should be output in CPR IPPV
mode in the event of rising or falling
expiratory minute volume. In CPR
30:2 and CPR 15:2 modes this alarm
is deactivated as a rule.
f
(Only with Flow
measurement + ASB
option)
☑
☐
Here you can determine whether an
alarm should be output in CPR IPPV
mode in the event of a rising
respiratory rate. In CPR 30:2 and CPR
15:2 modes this alarm is deactivated
as a rule.
Pause cont. 20 s - 60 s
Here you can set the maximum
duration of the ventilation pause for
the analysis phase of the defibrillator
during the continuous ventilation.
CPR CCSV
Hands-off time 5 s - 55 s
Here you can set when the hands-off
time alarm should appear.
Time until backup
vent.
10 s - 60 s
Here you can set when the back-up
ventilation should be initiated. The
hands-off time is always less than
the time to back-up ventilation.
f CCSV /
80/140
90/130
100/120
You can set the alarm limit for the
chest compression rate here.
f CCSV / alarm
☑
☐
Here you can determine whether or
not an alarm should be emitted if the
compression rate is too slow or too
fast.
Parameter Possible values Description