136 EN MEDUMAT Standard
2
WM 68071 05/2021
6 Operator menu
RSI mode (only with Flow measurement + ASB option or Capnography option)
View (only with
activated Capnography
option)
Pressure gauge
Pressure/CO
2
curve
Here you can choose between a
pressure gauge view and a curve
view for the RSI mode.
Contin. Mode (only with
activated Pressure
controlled ventilation
modes option)
IPPV
BiLevel + ASB
Here you can determine whether
the device switches to IPPV mode
or BiLevel + ASB mode after RSI
Manual.
MVe / (only with Flow
measurement + ASB
option)
☑
☐
Here you can determine whether
an alarm should be output in RSI
Demand mode in the event of rising
or falling expiratory minute volume.
In RSI Manual mode this alarm is
deactivated as a rule.
f (only with Flow
measurement + ASB
option)
☑
☐
Here you can determine whether
an alarm should be output in RSI
Demand mode in the event of a
rising respiratory rate. In RSI Manual
mode this alarm is deactivated as
a rule.
etCO
2
/ (only with
Capnography option)
☑
☐
Here you can determine whether an
alarm should be output in the event
of rising or dropping end-expiratory
CO
2
.
Apnea mode (only with activated Pressure controlled ventilation modes option)
IPPV
BiLevel + ASB
Here you can set the apnea
ventilation mode for the CPAP
and CPAP + ASB modes.
Height
Vt per kg bodyweight 4 ml/kg - 10 ml/kg
Here you can set the tidal volume in
milliliters per kilogram body weight.
In the process, a variable is used to
convert the height to a tidal volume
(see "14.2 Calculation of body
weight on the basis of body height",
page 258).