WM 68071 05/2021
MEDUMAT Standard
2
EN 137
6 Operator menu
Setting height*
Depending on the height selected (tidal volume Vt in ml per kg
bodyweight) the height which can be set is restricted to the
following minimum values:
PEEP Warning
1 mbar - 21 mbar
Here you can set a limit value for the
positive end-expiratory pressure.
A warning is then given on the
display if this value is reached or
exceeded. In this case, the PEEP
field in the bottom right of the
display turns red.
Parameter Possible values Description
Tidal volume Vt in ml per kg
bodyweight
Minimum height which can be
set in cm
490
580
670
765
860
955
10 50
For the smallest height which can be set, the tidal volume is always
at least 50 ml.