7 Description of the modes
7.4.2 RSI mode
The RSI mode supports you in the induction of anesthesia (TIVA).
It is used on all patients with an increased risk of a pulmonary
aspiration.
Following the selection of the RSI mode, the device launches the
oxygen demand function immediately for the pre-oxygenation of a
spontaneously breathing patient.
For intubation switch to the Manual function. The I:E ratio is
always 1:1. With the supplied MEDUtrigger, this function now
enables manual ventilation with a defined volume and a defined
pressure limitation. The Manual function can be used for checking
the position of the endotracheal tube or as a fallback option should
it prove difficult to secure the airway.
You cannot activate Air Mix in the Demand and Manual functions.
Risk of injury from deactivated alarms in the additional
functions CPR and RSI!
Through deactivation in the operator menu, no alarms for flow
and CO
2
monitoring are output in the additional functions CPR
and RSI. This can injure the patient.
Notify the user if the alarms in the additional functions CPR
and RSI have been deactivated.
Observe the alarm limits during ventilation in the additional
functions CPR and RSI.
Description
Abbreviation RSI
Long form Rapid Sequence Induction
Type Volume-controlled
Requirement None
Ventilation parameters
Left-hand navigation knob Vt (only with RSI Manual)
Central navigation knob -
Right-hand navigation knob
• Emergency mode
•RSI mode
• pMax