166 EN MEDUMAT Standard
2
WM 68071 05/2021
7 Description of the modes
Following successful airway management, switch to continuous
ventilation mode with the Contin. function. The device
automatically switches to one of the following modes with the
ventilation parameters preset for the patient group
(see "14.1.10 Factory settings for emergency modes and
ventilation modes", page 257):
• IPPV
• BiLevel + ASB (only with activated BiLevel + ASB option).
If the Capnography option is activated, you can set a pressure
gauge view or curve view in CPR mode (see "6.3.8 Presets
patient", page 132).
7.4.3 Demand mode
The Demand mode serves to (pre)oxygenate spontaneously
breathing patients via a ventilation mask. The patient must trigger
inspiration himself in Demand mode. If there is a FlowCheck sensor
this recognizes the respiratory effort, otherwise the underpressure
created is used. You can select the following operation in Demand
mode:
• Concentrator oxygen mode
• 100 % oxygen mode
Air Mix cannot be activated in Demand mode.
Description
Abbreviation -
Long form Demand
Type Pressure-controlled
Requirement Demand option is activated
Ventilation parameters
Left-hand navigation knob -
Central navigation knob -
Right-hand navigation knob
• pMax
• Emergency mode