180 EN MEDUMAT Standard
2
WM 68071 05/2021
8 Hygienic reprocessing
14. If available: Remove the protective cap strap from the patient
valve:
Position 1
or
Position 2 (only with reusable breathing circuits with flow
measurement and CO
2
measurement)
15. Disassemble the patient valve.
Result The reusable breathing circuit is disassembled.
12