74 EN MEDUMAT Standard
2
WM 68071 05/2021
4 Preparation and operation
4. If necessary: Set the parameters of the ventilation mode.
5. Select the start field.
Result The ventilation mode is changed.
4.7.6 Operating the device in oxygen or Air Mix mode
Requirement The device is switched on.
A ventilation mode is set.
1. Briefly press the Air Mix button .
Air Mix appears in the status line and the device is operated in
Air Mix mode.
2. Briefly press the Air Mix button :
•If 100 % O
2
is set as the supply gas (see "6.3.7 Device
configuration", page 129): 100 % O
2
appears in the status
line and the device is operated with 100 % oxygen.
or
•If 93 % O
2
is set as the supply gas (see "6.3.7 Device
configuration", page 129): appears in the status line
and the device is operated with concentrator oxygen.
93%
O2