MEDUMAT Standard
2
EN 73
4 Preparation and operation
Depending on the preset in the operator menu, the device
switches to one of the following modes with the ventilation
parameters preset for the patient group (see "14.1.10 Factory
settings for emergency modes and ventilation modes",
page 257) and shows a pressure gauge view:
•IPPV
• BiLevel + ASB (only with activated BiLevel + ASB option)
Result An emergency mode for a particular patient group is activated.
4.7.5 Changing the ventilation mode
Requirement • The device is switched on.
• A ventilation mode is set.
1. Briefly press the menu button .
The user menu opens.
2. Select the Mode field using the right-hand navigation knob.
3. Select a ventilation mode.
You can adjust the preset ventilation parameters for the
emergency modes in the operator menu: Operator menu |
Presets patient.