WM 68071 05/2021
MEDUMAT Standard
2
EN 79
4 Preparation and operation
or
Change to CCSV or IPPV (see "4.7.12 Changing
the ventilation mode in CPR mode", page 84) with the right
navigation knob.
Result Ventilation is carried out in CPR Manual mode.
4.7.10 Performing ventilation in CPR IPPV mode
Requirement • The device is switched on.
• CCSV option is deactivated.
1. Briefly press the CPR button .
Depending on the preset in the operator menu, the device
switches to either the CPR IPPV or the CPR Manual mode
(Operator menu | Presets patient | CPR mode | Start mode).
2. If is already set in the operator menu:
Continue ventilation in CPR IPPV mode.
or
If Manual is set in the operator menu: Switch to IPPV
with the right-hand navigation knob (see "4.7.12 Changing
the ventilation mode in CPR mode", page 84).
3. If the patient experiences spontaneous circulation again:
Cancel the CPR IPPV mode with the CPR button .
After exiting the CPR mode, the device always switches to
IPPV mode.
Delay in treatment due to simultaneous metronome outputs
from the ventilator and the defibrillator!
If the ventilator is used together with a defibrillator
(MEDUCORE Standard or MEDUCORE Standard
2
) which can also
emit a metronome sound, the simultaneous metronome outputs
from the defibrillator and the ventilator may confuse the user and
cause delays in treatment.
If the ventilator and defibrillator are being used at
the same time, switch off the metronome sound of
MEDUMAT Standard
2
.
Do not perform resuscitation without a metronome.