80 EN MEDUMAT Standard
2
WM 68071 05/2021
4 Preparation and operation
Result Ventilation is carried out in CPR IPPV mode.
Pausing ventilation in CPR IPPV mode
During the analysis of the defibrillator, you can pause ventilation
in order to avoid artifacts in the analysis.
Requirement The device is switched on.
The CPR mode is set.
Continuous ventilation IPPV is activated (CPR IPPV).
1. Turn or press the right-hand navigation knob.
2. Select the field.
The ventilation is paused for the interval time set in the
operator menu (max. 60 seconds). A countdown indicates
the remaining time.
After exiting the CPR mode, the device always switches to
IPPV mode.