CE CONFORMITY: This product is in conformity with the protection re-
quirements of the following European Council Directive: 89/336/EEC,the
EMC Directive and 73/23/EEC, Low Voltage Directive. Conformity of
this product with any other “CE Mark” Directive(s) shall not be assumed.
EMC Directive / Standard PC Conformity Notes
ELECTROMAGNETIC COMPATIBILITY:89/336/EEC,
EMC Directive
EMISSIONS: EN 50081-2-1993, Generic Emissions,
Industrial
EN 55011-1991, Group 1, Class A, Industrial Control 30~37 dBµV/m at 30m 1
Equipment, 150 KHz-1000 MHz
IMMUNITY: EN 50082-2-1995, Generic Immunity, PERFORMANCE: Unless otherwise noted, the
Industrial. Performance Criteria(PC) A or B as applicable. performance of this product, at the specified levels of
electromagnetic interference, is within the specifications
for “Performance Under Rated Conditions”.
EN 61000-4-2-1995 (IEC 1000-4-2-1995),ESD, B 4kV Contact
Electrostatic Discharge B 8kV Air
ENV 50140-1993 (IEC 1000-4-3-1995),Radiated RF A 10 V/m 1
Fields, 80-1000MHz
ENV 50204-1995 Radiated RF Fields, 900 MHz A 10 V/m 1
EN 61000-4-4-1995 (IEC 1000-4-4-1995), Electrical Fast B 2 kV Process Measurement & Control
Transients/Burst
ENV 50141-1993, Conducted RF Fields, 150 kHz-80 MHz A 10V 1.2
NOTES:
PC = Performance Criteria
1. Twist pair cables required for all I/O interface circuits.
In case of remote model two core double shield cable in metal conduit
pipe required for the input line in connection with detector.
2. Error up ±100% of the converter output value between the noise fre-
quency 1 to 80 MHz.
LV Directive Conformity
LOW VOLTAGE DIRECTIVE:73/23/EEC EN 61010-1, Safety requirements for electrical equipment for measurement,
control and laboratory use Part 1: General requirements
MagneW3000 PLUS Electromagnetic Flowmeter
CE Conformity Supplement