SY
Switches
between
the
normal
and
adjustment
modes/queries the current setting.
When normal
condition/adjustment
Syntax for setting
SYm<delimiter>
Syntax for query
SY?<delimiter> ⇒ Response: SYm<delimiter>
Description of parameter
m: Mode
m=0: Normal mode m=1: Adjustment mode
CD
Sets
the
sourced
setpoint/queries
the
current
setting.
When adjustment
Syntax for setting
CDm<delimiter>
Syntax for query
DC? <delimiter> ⇒ Response: DCm<delimiter>
Description of parameter
m: Sourced setpoint (8 digits) ex. +1.00003
CL
Sets
the
adjustment
item/queries
the
current
setting.
When adjustment
Syntax for setting
CLm<delimiter>
Syntax for query
CL?<delimiter> ⇒ Response: CLm<delimiter>
Description of parameter
m: Adjustment item
m=3: Adjustment of source m=4: Adjustment of measurement
CP Sets the adjustment point. When adjustment
Syntax for setting
CPm<delimiter>
Description of parameter
m: Adjustment point
m=0: FS adjustment m=1: Zero adjustment
CW Saves the adjusted data. When adjustment
Syntax for setting
CW<delimiter>
Be sure to execute CW command after adjustment for each function/range.
Without executing CW command, the adjusted value will be deleted when the
power is turned off.