EasyManua.ls Logo

YOKOGAWA CA71 HANDY CAL - Page 85

YOKOGAWA CA71 HANDY CAL
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IM CA71-E
8-9
Communication Function
1
2
3
4
5
6
7
8
9
10
11
12
13
App
SY
Switches
between
the
normal
and
adjustment
modes/queries the current setting.
When normal
condition/adjustment
Syntax for setting
SYm<delimiter>
Syntax for query
SY?<delimiter> Response: SYm<delimiter>
Description of parameter
m: Mode
m=0: Normal mode m=1: Adjustment mode
CD
Sets
the
sourced
setpoint/queries
the
current
setting.
When adjustment
Syntax for setting
CDm<delimiter>
Syntax for query
DC? <delimiter> Response: DCm<delimiter>
Description of parameter
m: Sourced setpoint (8 digits) ex. +1.00003
CL
Sets
the
adjustment
item/queries
the
current
setting.
When adjustment
Syntax for setting
CLm<delimiter>
Syntax for query
CL?<delimiter> Response: CLm<delimiter>
Description of parameter
m: Adjustment item
m=3: Adjustment of source m=4: Adjustment of measurement
CP Sets the adjustment point. When adjustment
Syntax for setting
CPm<delimiter>
Description of parameter
m: Adjustment point
m=0: FS adjustment m=1: Zero adjustment
CW Saves the adjusted data. When adjustment
Syntax for setting
CW<delimiter>
Be sure to execute CW command after adjustment for each function/range.
Without executing CW command, the adjusted value will be deleted when the
power is turned off.
8.7 Detailed Description of Commands

Table of Contents

Related product manuals