CMF Queries the measurement function. When adjustment
Syntax for query
CMF?<delimiter> ⇒ Response: CMFm<delimiter>
Description of parameter
m: Measurement function
m=0: AC 300V m=1: 100V m=2: 10V
m=3: 1V m=4: 100mV m=5: Resistance
m=6: Frequency m=7: Current
CSF Queries the source function. When adjustment
Syntax for query
CSF?<delimiter> ⇒ Response: CSFm<delimiter>
Description of parameter
m= Function
m=0: 30V m=1: 10V m=2: 1V m=3: 100mV
m=4: Resistance m=5: Pulse m=6: 20mA m=7: 4 to 20mA
m=8: 20mASINK
OM Queries memory data.
When normal
condition
Syntax for query
OMm<delimiter>
Response:
abcde, fghij [, fghij ] <delimiter>
<Header section of measured value>
a= V: Voltage A: Current O: Resistance T: Temperature F: Frequency
b= DC: Direct current AC: Alternating current
c= N: Normal O: Overrange E: No data
<Data section of measured value>
d = Measured value, mantissa part (7 digits)
e = Measured value exponent part (E- 3 / E+0 / E+3)
<Header section of sourced setpoint>
f= V: Voltage A: Current O: Resistance T: Temperature F: Frequency
g= DC: Direct current AC: Alternating current
h= N: Normal E: No data
i = Sourced setpoint, mantissa part (7 digits)
j = Sourced setpoint exponent part (E
- 3 / E+0 / E+3)
Description of parameter
m: Number of memory 1 to 50