EasyManua.ls Logo

YOKOGAWA DLM2000 Series - Page 101

YOKOGAWA DLM2000 Series
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-68
IM 710105-17E
:ANALysis:PANalyze<x>:SWLoss:SCALe:M
ODE
Function Sets or queries the scaling mode.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:MO
DE {AUTO|MANual}
:ANALysis:PANalyze<x>:SWLoss:SCALe:MO
DE?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:MO
DE AUTO
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:MO
DE?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:M
ODE AUTO
:ANALysis:PANalyze<x>:SWLoss:SCALe:S
ENSitivity
Function Sets or queries the span from the center value for
manual scaling.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:SE
NSitivity {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:SCALe:SE
NSitivity?
<x> = 1 or 2
<NRf>=−1.0000E+31to1.0000E+31
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:SENS
ITIVITY 10
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:SENS
ITIVITY?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:S
ENSITIVITY 10.0000E+00
:ANALysis:PANalyze<x>:SWLoss:TRANge
(Time Range)
Function Sets or queries the measurement time period.
Syntax
:ANALysis:PANalyze<x>:SWLoss:TRAN
ge {<NRf>,<NRf>}
:ANALysis:PANalyze<x>:SWLoss:TRANge?
<x> = 1 or 2
<NRf>=−5to5divisions(instepsof10divisions/
display record length)
Example
:ANALYSIS:PANALYZE1:SWLOSS:TRANGE−4,
4
:ANALYSIS:PANALYZE1:SWLOSS:TRANGE?
-> :ANALYSIS:PANALYZE1:SWLOSS:
TRANGE−4.00E+00,4.00E+00
:ANALysis:PANalyze<x>:SWLoss:RANGe
Function Sets or queries the measurement source window.
Syntax
:ANALysis:PANalyze<x>:SWLoss:RAN
Ge {MAIN|Z1|Z2}
:ANALysis:PANalyze<x>:SWLoss:RANGe?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:RANGE MAIN
:ANALYSIS:PANALYZE1:SWLOSS:RANGE?
-> :ANALYSIS:PANALYZE1:SWLOSS:RAN
GE MAIN
:ANALysis:PANalyze<x>:SWLoss:RDS
Function Sets or queries the on-resistance value for total
loss computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:R
DS {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:RDS?
<x> = 1 or 2
<NRf>=0to100(in1mΩsteps)
Example
:ANALYSIS:PANALYZE1:SWLOSS:RDS 1
:ANALYSIS:PANALYZE1:SWLOSS:RDS?
-> :ANALYSIS:PANALYZE1:SWLOSS:RDS 1.0
00E+00
:ANALysis:PANalyze<x>:SWLoss:SCALe?
Function Queries all scaling settings.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe?
<x> = 1 or 2
:ANALysis:PANalyze<x>:SWLoss:SCALe:C
ENTer
Function Sets or queries the center value for manual
scaling.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:CE
NTer {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:SCALe:CE
NTer?
<x> = 1 or 2
<NRf>=−1.0000E+31to1.0000E+31
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:CENT
ER 1
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:CENT
ER?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:C
ENTER 1.00000E+00
5.3 ANALysis Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Related product manuals