5-68
IM 710105-17E
:ANALysis:PANalyze<x>:SWLoss:SCALe:M
ODE
Function Sets or queries the scaling mode.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:MO
DE {AUTO|MANual}
:ANALysis:PANalyze<x>:SWLoss:SCALe:MO
DE?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:MO
DE AUTO
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:MO
DE?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:M
ODE AUTO
:ANALysis:PANalyze<x>:SWLoss:SCALe:S
ENSitivity
Function Sets or queries the span from the center value for
manual scaling.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:SE
NSitivity {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:SCALe:SE
NSitivity?
<x> = 1 or 2
<NRf>=−1.0000E+31to1.0000E+31
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:SENS
ITIVITY 10
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:SENS
ITIVITY?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:S
ENSITIVITY 10.0000E+00
:ANALysis:PANalyze<x>:SWLoss:TRANge
(Time Range)
Function Sets or queries the measurement time period.
Syntax
:ANALysis:PANalyze<x>:SWLoss:TRAN
ge {<NRf>,<NRf>}
:ANALysis:PANalyze<x>:SWLoss:TRANge?
<x> = 1 or 2
<NRf>=−5to5divisions(instepsof10divisions/
display record length)
Example
:ANALYSIS:PANALYZE1:SWLOSS:TRANGE−4,
4
:ANALYSIS:PANALYZE1:SWLOSS:TRANGE?
-> :ANALYSIS:PANALYZE1:SWLOSS:
TRANGE−4.00E+00,4.00E+00
:ANALysis:PANalyze<x>:SWLoss:RANGe
Function Sets or queries the measurement source window.
Syntax
:ANALysis:PANalyze<x>:SWLoss:RAN
Ge {MAIN|Z1|Z2}
:ANALysis:PANalyze<x>:SWLoss:RANGe?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:RANGE MAIN
:ANALYSIS:PANALYZE1:SWLOSS:RANGE?
-> :ANALYSIS:PANALYZE1:SWLOSS:RAN
GE MAIN
:ANALysis:PANalyze<x>:SWLoss:RDS
Function Sets or queries the on-resistance value for total
loss computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:R
DS {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:RDS?
<x> = 1 or 2
<NRf>=0to100(in1mΩsteps)
Example
:ANALYSIS:PANALYZE1:SWLOSS:RDS 1
:ANALYSIS:PANALYZE1:SWLOSS:RDS?
-> :ANALYSIS:PANALYZE1:SWLOSS:RDS 1.0
00E+00
:ANALysis:PANalyze<x>:SWLoss:SCALe?
Function Queries all scaling settings.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe?
<x> = 1 or 2
:ANALysis:PANalyze<x>:SWLoss:SCALe:C
ENTer
Function Sets or queries the center value for manual
scaling.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:CE
NTer {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:SCALe:CE
NTer?
<x> = 1 or 2
<NRf>=−1.0000E+31to1.0000E+31
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:CENT
ER 1
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:CENT
ER?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:C
ENTER 1.00000E+00
5.3 ANALysis Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals