5-69
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
:ANALysis:PANalyze<x>:SWLoss:ULEVel
Function Sets or queries the voltage level used to
determine the loss calculation period for total loss
computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:ULEV
el {<NRf>|<Voltage>}
:ANALysis:PANalyze<x>:SWLoss:ULEVel?
<x> = 1 or 2
<NRf>, <Voltage> = See the DLM2000 Features
Guide for this information.
Example
:ANALYSIS:PANALYZE1:SWLOSS:ULEVEL 1V
:ANALYSIS:PANALYZE1:SWLOSS:ULEVEL?
-> :ANALYSIS:PANALYZE1:SWLOSS:
ULEVEL 1.0000000E+00
:ANALysis:PANalyze<x>:SWLoss:UNIT
Function Sets or queries the power unit.
Syntax
:ANALysis:PANalyze<x>:SWLoss:UNIT {J|
WH}
:ANALysis:PANalyze<x>:SWLoss:UNIT?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:UNIT WH
:ANALYSIS:PANALYZE1:SWLOSS:UNIT?
-> :ANALYSIS:PANALYZE1:SWLOSS:UNIT WH
:ANALysis:PANalyze<x>:SWLoss:VCE
Function Sets or queries the collector-emitter saturation
voltage value for total loss computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:V
CE {<NRf>|<Voltage>}
:ANALysis:PANalyze<x>:SWLoss:VCE?
<x> = 1 or 2
<NRf>, <Voltage> = 0 to 50 V (in 100-mV steps)
Example
:ANALYSIS:PANALYZE1:SWLOSS:VCE 1V
:ANALYSIS:PANALYZE1:SWLOSS:VCE?
-> :ANALYSIS:PANALYZE1:SWLOSS:VCE 1.0
000000E+00
:ANALysis:PANalyze<x>:TYPE
Function Sets or queries the type of power supply analysis.
Syntax
:ANALysis:PANalyze<x>:TYPE {HARMonics
|I2T|OFF|SOA|SWLoss}
:ANALysis:PANalyze<x>:TYPE?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:TYPE HARMONICS
:ANALYSIS:PANALYZE1:TYPE?
-> :ANALYSIS:PANALYZE1:TYPE HARMONICS
:ANALysis:PMEAsure<x>?
Function Queries all power measurement settings.
Syntax
:ANALysis:PMEAsure<x>?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1?
-> :ANALYSIS:PMEASURE1
:ANALysis:PMEAsure<x>:IDPRoximal?
Function Queries all distal, mesial, and proximal settings.
Syntax
:ANALysis:PMEAsure<x>:IDPRoximal?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:IDPROXIMAL?
:ANALysis:PMEAsure<x>:IDPRoximal:MO
DE
Function Sets or queries the distal, mesial, and proximal
point mode setting.
Syntax
:ANALysis:PMEAsure<x>:IDPRoximal:MO
DE {PERCent|UNIT}
:ANALysis:PMEAsure<x>:IDPRoximal:MO
DE?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:IDPROXIMAL:MO
DE PERCENT
:ANALYSIS:PMEASURE1:IDPROXIMAL:MODE?
-> :ANALYSIS:PMEASURE1:IDPROXIMAL:MO
DE PERCENT
:ANALysis:PMEAsure<x>:IDPRoximal:PER
Cent
Function Sets or queries the distal, mesial, and proximal
points as percentages.
Syntax
:ANALysis:PMEAsure<x>:IDPRoximal:PERC
ent {<NRf>,<NRf>,<NRf>}
:ANALysis:PMEAsure<x>:IDPRoximal:PERC
ent?
<x> = 1 to 2
<NRf>, <NRf>, <NRf> = 0 to 100 (%, in steps of 1)
Example
:ANALYSIS:PMEASURE1:IDPROXIMAL:PERCE
NT 10,50,90
:ANALYSIS:PMEASURE1:IDPROXIMAL:PERCE
NT?
-> :ANALYSIS:PMEASURE1:IDPROXIMAL:PER
CENT 10,50,90
5.3 ANALysis Group