EasyManuals Logo
Home>YOKOGAWA>Test Equipment>DLM2000 Series

YOKOGAWA DLM2000 Series User Manual

YOKOGAWA DLM2000 Series
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
5-69
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
:ANALysis:PANalyze<x>:SWLoss:ULEVel
Function Sets or queries the voltage level used to
determine the loss calculation period for total loss
computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:ULEV
el {<NRf>|<Voltage>}
:ANALysis:PANalyze<x>:SWLoss:ULEVel?
<x> = 1 or 2
<NRf>, <Voltage> = See the DLM2000 Features
Guide for this information.
Example
:ANALYSIS:PANALYZE1:SWLOSS:ULEVEL 1V
:ANALYSIS:PANALYZE1:SWLOSS:ULEVEL?
-> :ANALYSIS:PANALYZE1:SWLOSS:
ULEVEL 1.0000000E+00
:ANALysis:PANalyze<x>:SWLoss:UNIT
Function Sets or queries the power unit.
Syntax
:ANALysis:PANalyze<x>:SWLoss:UNIT {J|
WH}
:ANALysis:PANalyze<x>:SWLoss:UNIT?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:UNIT WH
:ANALYSIS:PANALYZE1:SWLOSS:UNIT?
-> :ANALYSIS:PANALYZE1:SWLOSS:UNIT WH
:ANALysis:PANalyze<x>:SWLoss:VCE
Function Sets or queries the collector-emitter saturation
voltage value for total loss computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:V
CE {<NRf>|<Voltage>}
:ANALysis:PANalyze<x>:SWLoss:VCE?
<x> = 1 or 2
<NRf>, <Voltage> = 0 to 50 V (in 100-mV steps)
Example
:ANALYSIS:PANALYZE1:SWLOSS:VCE 1V
:ANALYSIS:PANALYZE1:SWLOSS:VCE?
-> :ANALYSIS:PANALYZE1:SWLOSS:VCE 1.0
000000E+00
:ANALysis:PANalyze<x>:TYPE
Function Sets or queries the type of power supply analysis.
Syntax
:ANALysis:PANalyze<x>:TYPE {HARMonics
|I2T|OFF|SOA|SWLoss}
:ANALysis:PANalyze<x>:TYPE?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:TYPE HARMONICS
:ANALYSIS:PANALYZE1:TYPE?
-> :ANALYSIS:PANALYZE1:TYPE HARMONICS
:ANALysis:PMEAsure<x>?
Function Queries all power measurement settings.
Syntax
:ANALysis:PMEAsure<x>?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1?
-> :ANALYSIS:PMEASURE1
:ANALysis:PMEAsure<x>:IDPRoximal?
Function Queries all distal, mesial, and proximal settings.
Syntax
:ANALysis:PMEAsure<x>:IDPRoximal?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:IDPROXIMAL?
:ANALysis:PMEAsure<x>:IDPRoximal:MO
DE
Function Sets or queries the distal, mesial, and proximal
point mode setting.
Syntax
:ANALysis:PMEAsure<x>:IDPRoximal:MO
DE {PERCent|UNIT}
:ANALysis:PMEAsure<x>:IDPRoximal:MO
DE?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:IDPROXIMAL:MO
DE PERCENT
:ANALYSIS:PMEASURE1:IDPROXIMAL:MODE?
-> :ANALYSIS:PMEASURE1:IDPROXIMAL:MO
DE PERCENT
:ANALysis:PMEAsure<x>:IDPRoximal:PER
Cent
Function Sets or queries the distal, mesial, and proximal
points as percentages.
Syntax
:ANALysis:PMEAsure<x>:IDPRoximal:PERC
ent {<NRf>,<NRf>,<NRf>}
:ANALysis:PMEAsure<x>:IDPRoximal:PERC
ent?
<x> = 1 to 2
<NRf>, <NRf>, <NRf> = 0 to 100 (%, in steps of 1)
Example
:ANALYSIS:PMEASURE1:IDPROXIMAL:PERCE
NT 10,50,90
:ANALYSIS:PMEASURE1:IDPROXIMAL:PERCE
NT?
-> :ANALYSIS:PMEASURE1:IDPROXIMAL:PER
CENT 10,50,90
5.3 ANALysis Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals