5-134
IM 710105-17E
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:
<Parameter>:{MAXimum|MEAN|MINimum|SD
EViation}?
Function Queries a statistical value of a waveform
parameter in Area2.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:<
Parameter>:{MAXimum|MEAN|MINimum|SDEV
iation}?
<x> of CHANnel<x> = 1 to 4 (1 or 2 for 2-channel
models)
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
<Parameter> = {AMPLitude|
AVERage|AVGFre
q|AVGPeriod|BWIDth|DELay|DT|DUTYcycl
e|ENUMber|FALL|FREQuency|HIGH|LOW|MA
Ximum|MINimum|NOVershoot|NWIDth|PERi
od|PNUMber|POVershoot|PTOPeak|PWIDth|
RISE|RMS|SDEViation|TY1Integ|TY2Integ
|V1|V2
}
Example (Below is an example for the average of CH1.)
:MEASURE:CHANNEL1:AREA2:AVERAGE:MAXIM
UM?
-> :MEASURE:CHANNEL1:AREA2:AVERAGE:MA
XIMUM 10.0000E+03
Description If the statistical value is immeasurable, the
DLM2000 returns “NAN” (not a number).
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:
<Parameter>:STATe
Function Sets or queries the on/off status of a waveform
parameter in Area2.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:<
Parameter>:STATe {<Boolean>}
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:<
Parameter>:STATe?
<x> of CHANnel<x> = 1 to 4 (1 or 2 for 2-channel
models)
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
<Parameter> = {AMPLitude|
AVERage|AVGFre
q|AVGPeriod|BWIDth|DELay|DT|DUTYcycl
e|ENUMber|FALL|FREQuency|HIGH|LOW|MA
Ximum|MINimum|NOVershoot|NWIDth|PERi
od|PNUMber|POVershoot|PTOPeak|PWIDth|
RISE|RMS|SDEViation|TY1Integ|TY2Integ
|V1|V2
}
Example (Below is an example for the average of CH1.)
:MEASURE:CHANNEL1:AREA2:AVERAGE:STA
TE ON
:MEASURE:CHANNEL1:AREA2:AVERAGE:STA
TE?
-> :MEASURE:CHANNEL1:AREA2:AVERAGE:ST
ATE 1
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:
<Parameter>:VALue?
Function Queries an automatically measured value of a
waveform parameter in Area2.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:<
Parameter>:VALue? [<NRf>[,STATus]]
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:<
Parameter>:VALue? [STATus]
<x> of CHANnel<x> = 1 to 4 (1 or 2 for 2-channel
models)
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
<Parameter> = {AMPLitude|
AVERage|AVGFreq
|AVGPeriod|BWIDth|DELay|DT|DUTYcycle|
ENUMber|FALL|FREQuency|HIGH|LOW|MAXim
um|MINimum|NOVershoot|NWIDth|PERiod|P
NUMber|POVershoot|PTOPeak|PWIDth|RISE
|RMS|SDEViation|TY1Integ|TY2Integ|V1
|V2
}
<NRf> = See the DLM2000 Features Guide for
this information.
Example (Below is an example for the average of CH1.)
:MEASURE:CHANNEL1:AREA2:AVERAGE:VAL
UE?
-> :MEASURE:CHANNEL1:AREA2:AVERAGE:VA
LUE 10.0000E+03
Description If the value is immeasurable, the DLM2000
returns “NAN” (not a number).
The <NRf> is used to specify which iteration of
automated measurement to query the measured
value from.
If <NRf> is set to 1, the oldest measured value
in the automated measurement memory is
queried.
If a measured value does not exist at the
specified iteration, the DLM2000 returns “NAN”
(not a number).
If <NRf> is omitted, the most recent measured
value is queried.
If cyclic statistical processing for automated
measurement is being executed and <NRf> is
specified, the measured values over a cycle in
iteration <NRf> from the left of the display on
the displayed waveform is queried. If <NRf>
is omitted, the measured values over the last
cycle on the displayed waveform is queried.
The last parameter “STATus,” which can be
omitted, is valid when the parameter is “FALL,”
“NWIDth,” “PERiod,” “PWIDth,” or “RISE.”
When “STATus” is omitted, the DLM2000
returns negative values of these parameters
when the measured values are less than the
measurement resolution. When “,STATus” is
included, the DLM2000 returns “LOW_RESOL.”
5.20 MEASure Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals